Hospital Breathing Machine Medical Ventilator Cardiopulmonary Resuscitation Ventilator
Technical parameters:
1. Host pressing depth: 30mm~65mm can be adjusted, each 5mm, the allowable error between the preset pressing depth and the actual pressing depth: ±2mm;
2. The default state of power-on: the compression ventilation ratio is 30:2 (continuous compression can be switched), the compression depth is 3cm, and the compression frequency is 100 times/min;
3. Compression frequency: 100 times/min, 110 times/min, 120 times/min;
4. Resuscitation tidal volume: 500ML cannot be adjusted;
5. Compression and release ratio: 1:1, non-belt type, to ensure no load on the chest;
6. Pressure safety valve: 0.6Mpa (≥0.6mpa automatic pressure relief); tolerance ±0.1Mpa; the gas circuit should withstand pressure not less than 0.8Mpa;
7. Pressing head pressure when the whole machine is working: ≤800±50N;
8. Input power: DC: ≤10W; AC: ≤50W;
9. The fastening fastness of the fixed belt is not less than 100N;
10. Insufficient air pressure alarms, the pressure depth is automatically reduced, and the required pressure depth is manually restored after the air pressure is restored;
11. Power display, low voltage alarm, start pause function;
12. Chest thickness measurement indication function;
13. The chest thickness of the applicable patient ranges from 155mm to 305mm;
14. Shock resistance, open swing arm classic vertical pressing;
15. Anti-fall, the operation panel is located above the patient;
16. The enclosure protection grade is IP55 (dust and water spray);
17. Transportation and storage requirements: a) Ambient temperature range: -20ºC~55ºC;
b)Relative humidity range: 30%~75%;
c) Atmospheric pressure range: 700hPa~1060hPa.
18. The control panel adopts the patented technology of LGF backlight, LCD screen display, suitable for operation under various light conditions;
19. The single-arm open and rotatable structure design is convenient for the application of other first-aid equipment and improves the rescue efficiency;
20. The operation panel is located above the patient to avoid contamination of vomit during rescue.
21. The host is equipped with a portable backpack, which is convenient for transportation and carrying.